User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Trig-Tek 203M:Charge Amplifier DATE: 13-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 36 NUMBER OF LINES: 285 CONFIGURATION: Fluke 5700A CONFIGURATION: HP 34401A STANDARD: DTB 1000pF ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD DTB 1000pF 1.004 HEAD INITIAL CONDITIONS 1.005 HEAD {} 1.006 JMP 2.001 1.007 EVAL 2.001 DISP Set the UUT as follows: 2.001 DISP CAL/OPER .............................. OPER 2.001 DISP SENSITIVITY switch .................... 1-11 2.001 DISP SENSITIVITY dial ...................... 10.0 2.001 DISP PEAK/RMS .............................. RMS 2.001 DISP SE/DIFF ............................... SE 2.002 DISP Set the UUT as follows: 2.002 DISP UNITS ................................. g'S 2.002 DISP FULL SCALE ............................ 100 2.002 DISP INPUT ................................. pC/g 2.002 DISP OUTPUT ................................ 10 mV/g 2.002 DISP LP FILTER ............................. MED 2.003 DISP Connect the 5700 to the ACCEL INPUT (top) 2.003 DISP connector on the Test card through the 1000pF standard 2.003 DISP capacitor. 2.004 DISP Connect the HP 34401A HI/LO input to the UUT 2.004 DISP ACCEL OUT BNC connector (2nd from bottom). 2.005 HEAD {* Acceleration Output *} 2.006 5700 0.70700V 139.9H S 2W 2.007 ACC 0.70700V 0.001U 2.008 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 2.009 MEME 2.010 MEMC V 3.000% 10mV/g #! Test Tol 0.02121, Sys Tol 0.0006535, TUR 32.4560 (>= 4.00). 3.001 DISP Set the UUT as follows: 3.001 DISP OUTPUT ................................ 100 mV/g 3.002 5700 0.707V 139.9H S 2W 3.003 ACC 7.0700V 0.001U 3.004 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 3.005 MEME 3.006 MEMC V 3.000% 100mV/g #! Test Tol 0.2121, Sys Tol 0.006535, TUR 32.4560 (>= 4.00). 4.001 HEAD {} 4.002 HEAD {* Meter Tests *} 4.003 JMP 5.001 4.004 EVAL 5.001 DISP Set the UUT as follows: 5.001 DISP OUTPUT ................................ 10 mV/g 5.002 5700 0.70700V 139.9H S 2W 5.003 MEMI Enter the UUT meter reading. 5.004 ACC 70.7g's 0.1U 5.005 MEME 5.006 MEMC g's 2.0U rms #! Test Tol 2, Sys Tol 0.1, TUR 20.0000 (>= 4.00). 6.001 5700 0.70700V 139.9H S 2W 6.002 DISP Set the UUT as follows: 6.002 DISP PEAK/RMS .............................. PEAK 6.003 MEMI Enter the UUT meter reading. 6.004 ACC 100.0g's 0.1U 6.005 MEME 6.006 MEMC g's 2.0U peak #! Test Tol 2, Sys Tol 0.1, TUR 20.0000 (>= 4.00). 7.001 HEAD {} 7.002 HEAD {* Range Tests *} 7.003 JMP 8.001 7.004 EVAL 8.001 DISP Set the UUT controls as follows: 8.001 DISP FULL SCALE ............................ 1000 8.002 5700 7.0700V 139.9H S 2W 8.003 ACC 1000g's 1U 8.004 MEMI Enter the UUT meter reading. 8.005 MEME 8.006 MEMC g's 20U #! Test Tol 20, Sys Tol 1, TUR 20.0000 (>= 4.00). 9.001 5700 7.0700V 139.9H S 2W 9.002 ACC 7.0700V 0.001U 9.003 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 9.004 MEME 9.005 MEMC V 3.000% 139.9H # 9.002 34401 7.0700V 3.000% 139.9H 2W #! Test Tol 0.2121, Sys Tol 0.006535, TUR 32.4560 (>= 4.00). 10.001 HEAD {}* Range Tests * 10.002 JMP 11.001 10.003 EVAL 11.001 DISP Set the UUT controls as follows: 11.001 DISP FULL SCALE ............................ 100 11.002 5700 0.707000V 139.9H S 2W 11.003 ACC 100.0g's 0.01U 11.004 MEMI Enter the UUT meter reading. 11.005 MEME 11.006 MEMC g's 2.0U #! Test Tol 2, Sys Tol 0.01, TUR 200.000 (>= 4.00). 12.001 5700 0.707000V 139.9H S 2W 12.002 ACC .70700V 0.001U 12.003 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 12.004 MEME 12.005 MEMC V 3.000% 139.9H #12.002 34401 0.70700V 3.000% 139.9H 2W #! Test Tol 0.02121, Sys Tol 0.0006535, TUR 32.4560 (>= 4.00). 13.001 HEAD {}* Range Tests * 13.002 JMP 14.001 13.003 EVAL 14.001 DISP Set the UUT controls as follows: 14.001 DISP FULL SCALE ............................ 10 14.002 5700 70.700mV 139.9H S 2W 14.003 ACC 10.00g's 0.01U 14.004 MEMI Enter the UUT meter reading. 14.005 MEME 14.006 MEMC g's 0.2U #! Test Tol 0.2, Sys Tol 0.01, TUR 20.0000 (>= 4.00). 15.001 5700 70.700mV 139.9H S 2W 15.002 ACC 70.700mV 0.001U 15.003 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 15.004 MEM* 1000 15.005 MEME 15.006 MEMC mV 3.000% 139.9H #16.001 34401 70.700mV 3.000% 139.9H 2W #! Test Tol 0.002121, Sys Tol 7.535e-005, TUR 28.1486 (>= 4.00). 16.001 HEAD {} 16.002 HEAD {* DC Output *} 16.003 JMP 17.001 16.004 EVAL 17.001 5700 70.700mV 139.9H S 2W 17.002 DISP Connect the HP 34401A to the UUT DC OUTPUT BNC. 17.003 34401 10.000V 0.200U 2W #! Test Tol 0.2, Sys Tol 0.008, TUR 25.0000 (>= 4.00). 18.001 HEAD {} 18.002 HEAD {* Displacement Output Tests *} 18.003 JMP 19.001 18.004 EVAL 19.001 DISP Set the UUT controls as follows: 19.001 DISP UNITS ................................. MILS 19.001 DISP FULL SCALE ............................ 100 19.002 5700 0.70700V 139.9H S 2W 19.003 DISP Connect the HP 34401A to the UUT DISP OUT BNC. 19.004 ACC 100mils 0.1U 19.005 MEMI Enter the UUT meter reading. 19.006 MEME 19.007 MEMC mils 3U #! Test Tol 3, Sys Tol 0.1, TUR 30.0000 (>= 4.00). 20.001 5700 0.70700V 139.9H S 2W 20.002 ACC 1.0000V 0.001U 20.003 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 20.004 MEME 20.005 MEMC V 0.0300U 139.9H #20.002 34401 1.0000V 0.0300U 139.9H 2W #! Test Tol 0.03, Sys Tol 0.0008, TUR 37.5000 (>= 4.00). 21.001 HEAD {} 21.002 HEAD {* SENSITIVITY Dial Test *} 21.003 JMP 23.003 21.004 EVAL 22.001 JMP 23.001 22.002 EVAL 23.001 DISP Reset the UUT SENSITIVITY dial to 10.0 23.002 JMP 23.005 23.003 DISP Set the UUT controls as follows: 23.003 DISP UNITS ................................. g's 23.004 DISP Connect the HP 34401A to the UUT ACCEL OUT BNC. 23.005 5700 0.70700V 139.9H S 2W 23.006 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 23.007 MEM* 1000 23.008 MEME >1 23.009 5700 70.700mV 139.9H S 2W 23.010 DISP Set the UUT controls as follows: 23.010 DISP SENSITIVITY dial ...................... 1.0 23.011 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 23.012 MEM* 1000 23.013 MEME 23.014 MEME <1 23.015 MEMC mV 3% #! Test Tol 0.002121, Sys Tol 1.707e-005, TUR 124.253 (>= 4.00). 24.001 HEAD {} 24.002 HEAD {* SENSITIVITY Range Switch Test *} 24.003 JMP 26.001 24.004 EVAL 25.001 DISP Reset the UUT SENSITIVITY range switch to 1-11. 25.002 JMP 26.002 25.003 EVAL 26.001 DISP Set the UUT controls as follows: 26.001 DISP SENSITIVITY dial ...................... 10.0 26.002 5700 0.70700V 139.9H S 2W 26.003 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 26.004 MEM* 1000 26.005 MEME >1 26.006 DISP Set the UUT controls as follows: 26.006 DISP SENSITIVITY switch .................... 10-110 26.007 5700 7.0700V 139.9H S 2W 26.008 IEEE [@34401][D1000]MEAS:VOLT:AC?[D2000][I] 26.009 MEM* 1000 26.010 MEME 26.011 MEME <1 26.012 MEMC mV 3% 27.001 HEAD {} 27.002 HEAD {* LP FILTER Tests *} 27.003 HEAD {HIGH} 27.004 HEAD * LP FILTER Tests * 27.005 JMP 28.001 27.006 EVAL 28.001 DISP Set the UUT controls as follows: 28.001 DISP LP FILTER ............................. HIGH 28.001 DISP SENSITIVITY switch .................... 1-11 28.002 5700 0.707V 139.9H S 2W 28.003 DISP Adjust the UUT SENSITIVITY control for a UUT meter 28.003 DISP reading of 100.0. 28.004 MESS Adjust the Stimulus for a reading of 70.7 on the 28.004 MESS UUT meter. 28.005 5700 30.00kH 10.0% V 2W #! T.U.R. not calculated because MOD1 value not specified. #27.004 5700 27.00kH V S 2W #27.005 MEMI Enter the UUT meter reading. #27.006 MATH mem1 = 70.7 #27.007 MEME #27.008 MEMC g's 1.0U HIGH 29.001 HEAD {} 29.002 HEAD {MED} 29.003 HEAD * LP FILTER Tests * 29.004 JMP 30.001 29.005 EVAL 30.001 DISP Set the UUT controls as follows: 30.001 DISP LP FILTER ............................. MED 30.002 5700 0.707V 139.9H S 2W 30.003 DISP Adjust the UUT SENSITIVITY control for a UUT meter 30.003 DISP reading of 100.0. 30.004 MESS Adjust the Stimulus for a reading of 70.7 on the 30.004 MESS UUT meter. 30.005 5700 10.000kH 10.0% V 2W #! T.U.R. not calculated because MOD1 value not specified. #29.004 5700 9.70kH V S 2W #29.005 MEMI Enter the UUT meter reading. #29.006 MATH mem1 = 70.7 #29.007 MEME #29.008 MEMC g's 1.0U MED 31.001 HEAD {} 31.002 HEAD {LOW} 31.003 HEAD * LP FILTER Tests * 31.004 JMP 32.001 31.005 EVAL 32.001 DISP Set the UUT controls as follows: 32.001 DISP LP FILTER ............................. LOW 32.002 5700 0.707V 139.9H S 2W 32.003 DISP Adjust the UUT SENSITIVITY control for a UUT meter 32.003 DISP reading of 100.0. 32.004 MESS Adjust the Stimulus for a reading of 70.7 on the 32.004 MESS UUT meter. 32.005 5700 3.000kH 10.0% V 2W #! T.U.R. not calculated because MOD1 value not specified. #31.004 5700 2.70kH V S 2W #31.005 MEMI Enter the UUT meter reading. #31.006 MATH mem1 = 70.7 #31.007 MEME #31.008 MEMC g's 1.0U LOW 33.001 HEAD {} 33.002 HEAD {* CAL Tests *} 33.003 JMP 34.001 33.004 EVAL 34.001 DISP Set the UUT controls as follows: 34.001 DISP OPER/CAL .............................. CAL 34.002 DISP Connect the HP 34401A to the UUT front panel 34.002 DISP INPUT/MONITOR BNC. 34.003 HEAD 34.004 IEEE [@34401]MEAS:VOLT:AC?[D2000][I] 34.005 ACC 1.0000V 0.001U 34.006 MEME 34.007 MEMC V 1% #! Test Tol 0.01, Sys Tol 0.0008, TUR 12.5000 (>= 4.00). 35.001 34401 139.9H 0.8U 1V 2W #! Test Tol 0.8, Sys Tol 0.01399, TUR 57.1837 (>= 4.00). 36.001 HEAD {} 36.002 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 3 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.